Users Manual
54
The Meter allocates memory in such a way as to guarantee
capturing all of the user-specified sample intervals. Event
records will also be captured until the Meter detects the allocated
memory has been used up. At this point, events are not
recorded, but the event counter continues to advance to indicate
the total number of events that occurred. A plus (+) sign appears
after the event count to indicate this condition.
Note
The maximum number of recorded sample intervals is
10,000. The maximum number of recorded events is
15,000 minus the number of sample intervals. These
maximum numbers are proportionally decreased when
the available memory is low.
To change either of the two recording variables, use the cursor
buttons to position the menu selector next to the desired menu
item and press the softkey labeled Edit. Use 5 6 7 and 8 to
move between and set each digit of the selected variable.
If the battery level is anything but full, a message appears at the
bottom of the record menu reminding you of the battery level
before you start the recording session.
Setting the Event Threshold Value
Press the softkey labeled Setup to access the setup menu.
Using the cursor buttons, move the menu selector next to the
menu item labeled Recording and press the softkey labeled
Recording to open the recording setup screen. Using the cursor
buttons, move the menu selector next to the menu item labeled
Event Threshold for Recording and then press the softkey
labeled Edit. Press 7 or 8 to scroll through the event threshold
values. With the desired value selected, press the softkey
labeled Close.
Starting a Recording Session
Once the variables are set, press the softkey labeled Start, at
which point s appears in the display and the green LED
surrounding the power button (O) flashes. Table 8 shows the
recording display and describes the information displayed.
The Menu, Setup, Reference, and Temperature Offset softkey
functions are not available while the Meter is recording. This
ensures that measurements within a recording session are
consistent.
To extend battery life while recording, the Meter may enter a
battery-saver mode five minutes after a push-button is pressed
or IR communications activity ends. If the Auto Power-Off
timeout is set to never, the battery-saver mode is disabled.
Stopping a Recording Session
The recording session will continue until the allocated memory is
used, the batteries expire, the rotary switch is moved, a probe is
inserted or removed from the A or mA/μA jacks, or the session is
terminated by pressing the softkey labeled Stop.
Table 9 shows the display and describes the displayed
information after stopping a recording session.
After stopping a recording session, choose to save the recording
session, view the trend data (see the “Viewing Trend Data”
section), or close the recording session. If the session wasn’t
saved before pressing the Close softkey, the data is lost.
64

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Fluke 289 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Fluke 289

Fluke 289 Bedienungsanleitung - Deutsch - 92 seiten

Fluke 289 Bedienungsanleitung - Holländisch - 90 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info