783864
2
Zoom out
Zoom in
Vorherige Seite
1/2
Nächste Seite
11. The monitor requires 6 hours to cool from the maximum storage tempera-
ture between uses until the monitor is ready for its INTENDED USE when
the ambient temperature is 20°C.
12. Not servicing/maintenance while the monitor is in use.
EXPLANATION OF SYMBOLS ON UNIT
7. Displaying stored results
Picture 8-1 Picture 8-2 Picture 8-3
Picture 8-4 Picture 8-5 Picture 8-6
a. In Standby Mode, press “MEM” button to display the stored results. The cur-
rent memory bank will blink and the amount of results in this bank will be
displayed. Press “START” button to change over to other bank. Conrm your
selection by pressing “MEM” button. The current bank can also be conrmed
automatically after 5 seconds with no operation. See picture 8-1.
b. After selecting the memory bank, the LCD will display the average values in
this bank. See picture 8-2. If no result stored, LCD will show zeros. See picture
8-5.
c. Press “MEM” button, LCD will display the average value of all the results
which is measured from 5 o’clock to 9 o’clock in last 7 days in the current
user memory zone. See picture 8-3. If no result stored from 5 o’clock to 9
o’clock in last 7 days, LCD will display “0” for blood pressure and pulse rate.
d. Press “MEM” button again, LCD will display the average value of all the results
which is measured from 18 o’clock to 20 o’clock in last 7 days in the current
user memory zone. If no result stored from 18 o’clock to 20 o’clock in last 7
days, LCD will display “0” for blood pressure and pulse rate. See picture 8-4.
e. Then press the “MEM” button, the most recent result will be displayed. See
picture 8-6. Followed by, the blood pressure and pulse rate will be shown
separately. Irregular heartbeat symbol (if any) will blink. Press “MEM” button
again to review the next result. In this way, repeatedly pressing the “MEM”
button displays the respective results measured previously. If no result stored,
LCD will show zeros. See picture 8-5.
f. When displaying the stored results, the monitor will turn off automatically after
1 minute of no operation. You can also press the button “START” to turn off
the monitor manually.
8. Deleting measurements from the memory
When any result is displaying, keeping on pressing button “MEM” for three sec-
onds, all results will be deleted. Press the button “MEM” or “START”, the monitor
will turn off.
Picture 9-1 Picture 9-2
9. Assessing high blood pressure for adults
The following guidelines for assessing high blood pressure (without regard to
age or gender) have been established by the World Health Organization (WHO).
Please note that other factors (e.g. diabetes, obesity, smoking, etc.) need to be
taken into consideration. Consult with your physician for accurate assessment,
and never change your treatment by yourself.
12. Troubleshooting (2)
Classication of blood pressure for adults
Note: It is not intended to provide a basis of any type of rush toward emergency
conditions/diagnosis based on the color scheme and that the color scheme is
meant only to discriminate between the different levels of blood pressure.
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1
Emission
3. If this monitor is stored near freezing, allow it to acclimate to room tempera-
ture before use.
4. Do not attempt to disassemble this monitor.
5. It is recommended the performance should be checked every 2 years or after
repair. Please contact the service center.
6. Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after
moistened with water, diluted disinfectant alcohol, or diluted detergent.
7. No component can be maintained by user in the monitor. The circuit dia-
grams, component part lists, descriptions, calibration instructions, or other
information which will assist the user’s appropriately qualied technical per-
sonnel to repair those parts of equipment which are designated repairably
can be supplied.
8. The monitor can maintain the safety and performance characteristics for a
minimum of 10,000 measurements or three years, and the cuff integrity is
maintained after 1,000 open–close cycles of the closure.
9. It is recommended the cuff should be disinfected 2 times every week if need-
ed (For example, in hospital or in clinique). Wipe the inner side (the side
contacts skin) of the cuff by a soft cloth squeezed after moistened with Ethyl
alcohol (75-90%), then dry the cuff by airing.
10. The monitor requires 6 hours to warm from the minimum storage tempera-
ture between uses until the monitor is ready for its INTENDED USE when the
ambient temperature is 20 °C.
b. Then the current memory bank (U1, U2, U3 or U4) is shown. See picture 7-2.
Press “MEM” button to change over to other bank. Conrm your selection by
pressing “START” button. The current bank can also be
conrmed automatically after 5 seconds with no operation.
c.
If the monitor has stored results, the LCD will momentarily display the most re-
cent one. If no result has been stored, zero will appear on LCD. See Picture 7-3.
d. If the voice function is switched on, the monitor will speak out measurement
tips.
e. Then the monitor starts to seek zero pressure. See picture 7-4.
f. Then the monitor inates the cuff until sufcient pressure has built up for a
measurement. Then the monitor slowly releases air from the cuff and carries
out the measurement. Finally the blood pressure and pulse rate will be c a l -
culated and displayed on the LCD screen. The blood pressure classication
indicator and Irregular heartbeat symbol (if any) will blink on the screen. The
result will be automatically stored in the monitor. See Picture 7-5.
g. After measurement, the monitor will turn off automatically after 1 minute of
no operation. Alternatively, you can press the “START” button to turn off the
monitor manually.
h. During measurement, you can press the “START” button to turn off the moni-
tor manually.
Note: Please consult a health care professional for interpretation of pressure
measurements.
M32777-GB-Rev.3-02.20
21
22
23
24
20
19
27
28
29
30
26
25
33
34
35
36
32
31
32777 / KD-5920
PROBLEM POSSIBLE CAUSE SOLUTION
LCD Display shows battery
symbol
Low Battery Change the batteries
LCD shows “Er 0” Pressure system is unstable before
measurement
Don’t move and try again
LCD shows “Er 1” Fail to detect systolic pressure
LCD shows “Er 2” Fail to detect diastolic pressure
PROBLEM POSSIBLE CAUSE SOLUTION
LCD shows “Er 3” Pneumatic system blocked or cuff is
too tight during ination Connect the cuff correctly and
try again. If the monitor is still
abnormal, please contact the
local distributor or the factory
LCD shows “Er 4” Pneumatic system leakage or cuff is
too loose during ination
LCD shows “Er 5” Cuff pressure above 300mmHg
Measure again after ve
minutes. If the monitor is still
abnormal, please contact the
local distributor or the factory
LCD shows “Er 6” More than 160 seconds with cuff
pressure above 15 mmHg
LCD shows “Er 7” Inner memory error
LCD shows “Er 8” Device parameter checking error
LCD shows “Er A” Pressure sensor parameter error
No response when you
press button or load battery
Incorrect operation or strong
electromagnetic interference
Take out batteries for ve
minutes, and then reinstall
all batteries
Table 2
Enclosure Port
Table 3
Proximity elds from RF wireless communications equipment
10. Technical alarm description
The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with no delay if
the determined blood pressure (systolic or diastolic) is outside the rated range
specied in part SPECIFICACIONS. In this case, you should consult a physician
or check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory
and cannot be adjusted or inactivated. This alarm condition is assigned as low
priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on
LCD will disappear automatically after about 8 seconds.
PROBLEM POSSIBLE CAUSE SOLUTION
LCD Display shows
abnormal result
The cuff position was not correct or it
was not properly tightened
Apply the cuff correctly and try again
Body posture was not correct during
testing
Review the “BODY POSTURE DUR-
ING MEASUREMENT” sections of
the instructions and re-test
Speaking, arm or body movement, an-
gry, excited or nervous during testing
Re-test when calm and without
speaking or moving during the test
Irregular heartbeat
(arrhythmia)
It is inappropriate for people with
serious arrhythmia to use this Elec-
tronic Sphygmomanometer
Disposal: The product must not be disposed of along with other domestic
waste. The users must dispose of this equipment by bringing it to a specic
recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
WHO/ISH Denitions and classication of blood pressure levels
80 85 90 100 110
120
130
140
160
180
Severe Hypertension
Moderate Hypertension
Mild Hypertension
High-normal BP
Normal BP
Diastolic
(mmHg)
Sistolic
(mmHg)
BLOOD PRESSURE
CLASSIFICATION
SBP
mmHg
DBP
mmHg
Color
Indicator
Optimal
<120 <80 green
Normal
120-129 80-84 green
High-Normal
130-139 85-89 green
Grade 1 Hypertension 140-159 90-99 yellow
Grade 2 Hypertension 160-179 100-109 orange
Grade 3 Hypertension ≥180 ≥110 red
ANDON HEALTH CO., LTD.
No. 3 JinPing Street, YaAn Road, Nankai District,
Tianjin 300190, China
Made in China
iHealthLabs Europe SAS
36 Rue de Ponthieu,
75008, Paris, France
11. Troubleshooting (1)
MAINTENANCE
1. Do not drop this monitor or subject it to strong impact.
2. Avoid high temperature and solarization. Do not immerse the monitor
in water as this will result in damage to the monitor.
Follow instructions for use
Caution: read instructions (warnings) carefully
Type BF applied part
WEEE disposal
0197
Medical Device complies with Directive 93/42/EEC
Manufacturer
Date of manufacture
Authorized representative in the European commu-
nity
Serial number
IP20 Covering Protection rate
Keep in a cool, dry place
Keep away from sunlight
Product code
Lot number
Phenomenon Compliance Electromagnetic
environment
RF emissions CISPR 11 Group 1,
Class B
Home healthcare
environment
Harmonic distortion IEC 61000-3-2
Class A
Home healthcare
environment
Voltage uctuations
and icker
IEC 61000-3-3
Compliance
Home healthcare
environment
Phenomenon Basic EMC
standard
Immunity test levels
Home Healthcare
Environment
Electrostatic Discharge IEC 61000-4-2 ±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM eld IEC 61000-4-3 10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity elds from RF
wireless communications
equipment
IEC 61000-4-3 Refer to table 3
Rated power frequency
magnetic elds
IEC 61000-4-8 30A/m
50Hz or 60Hz
Test frequency
(MHz)
Band (MHz) Immunity test levels
Professional healthcare facility
environment
385 380-390 Pulse modulation 18Hz, 27V/m
450 430-470 FM, ±5kHz deviation, 1kHz sine,
28V/m
710 704-787 Pulse modulation 217Hz, 9V/m
745
780
810 800-960 Pulse modulation 18Hz, 28V/m
870
930
1720 1700-1990 Pulse modulation 217Hz, 28V/m
1845
1970
2450 2400-2570 Pulse modulation 217Hz, 28V/m
5240 5100-5800 Pulse modulation 217Hz, 9V/m
5500
5785
0197
IP20
2

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gima Jolly wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Gima Jolly

Gima Jolly Bedienungsanleitung - Deutsch - 2 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info