590606
26
Zoom out
Zoom in
Vorherige Seite
1/28
Nächste Seite
26 GMC-I Messtechnik GmbH
14.2 Fuses
The fuse is located at the measuring input. When the fuse is
defective, the measurements in the mΩ/Ω/ / ranges are
faulty. The error in the V range amounts to only 10%.
If a fuse should blow, eliminate the cause of overload before
placing the instrument back into service!
Inspecting the Integrated Fuse
Ð Set the rotary switch to the Ω measuring function.
Ð Short circuit the and Ω jacks.
Display < 0.2 Ω: fuse OK
Display approx. 200 kΩ: fuse blown or poor contact at
the fuse holder
Replacing the Fuse
Ð Open the instrument as described under “Replacing the
Storage Batteries” .
Ð Remove the blown fuse with the help of an abject such
as a test probe, and replace it with a new fuse.
Table of Allowable Fuses
* These fuses are available in packages of ten from our sales offices and
distributors.
Attention!
!
Use specified fuses only!
If fuses with other blowing characteristics, other
current ratings or other breaking capacities are used,
the operator is placed in danger, and protective
diodes, resistors and other components may be
damaged.
The use of repaired fuses and short-circuiting the
fuse holder are prohibited.
14.3 Housing
No special maintenance is required for the housing. Keep
outside surfaces clean. Use a slightly dampened cloth for
cleaning. Avoid the use of cleansers, abrasives and solvents.
Disposing of Batteries or Storage Batteries
Dispose of old batteries or storage batteries properly, i.e. at a
designated collection point.
14.4 Device Return and Environmentally Compatible Disposal
The METRA HIT27 is a category 9 product (monitoring and
control instrument) in accordance with ElektroG (German
Electrical and Electronic Device Law). This device is not
subject to the RoHS directive.
We identify our electrical and electronic devices (as
of August 2005) in accordance with WEEE 2002/
96/EG and ElektroG with the symbol shown to the
right per DIN EN 50419 .
These devices may not be disposed with the trash. Please
contact our service department regarding the return of old
devices (address see chapter 17).
If you use batteries or rechargeable batteries in your instrument
or accessories which no longer function properly, they must
be duly disposed of in compliance with the applicable
national regulations.
Batteries or rechargeable batteries may contain harmful
substances or heavy metal such as lead (PB), cadmium (CD)
or mercury (Hg).
They symbol shown to the right indicates that
batteries or rechargeable batteries may not be
disposed of with the trash, but must be delivered to
collection points specially provided for this purpose.
15 Multimeter Messages
The following messages appear at the main display, or the
auxiliary displays as required. See “Symbols used in the
Digital Display” on page 2 for messages displayed at visible
segments.
Blinking Unit of Measure
All measuring functions are balanced for each METRA HIT27
at the factory in accordance with the respective technical
specification. If a unit of measure blinks, this indicates that
balancing constants which have been established and saved
to the multimeter are no longer available for the respective
function. If this is the case, measurement results may deviate
from the specification. We recommend sending the
instrument to our Repair and Replacement Parts department
for rebalancing (see chapter 17).
Type Dimensions Article Number
FF (UR) 1.6 A/1000 V AC/DC (10 kA) 6.3 mm x 32 mm Z109C *
Message Function Meaning
0. L
Measuring Indicates overflow
leads open
4-wire-mΩ
Measuring current interrupted or fuse defective.
Pb Cd Hg
26

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln
1

Forum

gossen-metrawatt-metra-hit-27i

Suche zurücksetzen

  • Gibt es für das Gossen Metrawatt M205A einen Schaltplan? Eingereicht am 13-12-2019 10:46

    Antworten Frage melden

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt METRA HIT 27I wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Gossen Metrawatt METRA HIT 27I

Gossen Metrawatt METRA HIT 27I Kurzanleitung - Deutsch, Englisch - 4 seiten

Gossen Metrawatt METRA HIT 27I Bedienungsanleitung - Holländisch - 28 seiten

Gossen Metrawatt METRA HIT 27I Kurzanleitung - Französisch - 4 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info