Digital Automatic Blood Pressure Monitor
Model M3
Instruction Manual
English
Français
Deutsch
Italiano
Español
Nederlands
IM-HEM-7200-E-05-11/2010
5329491-4C
Manufacturer
OMRON HEALTHCARE Co., Ltd.
24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto
615-0084 JAPAN
EU-representative
OMRON HEALTHCARE EUROPE B.V.
Kruisweg 577, 2132 NA Hoofddorp THE NETHERLANDS
www.omron-healthcare.com
Production facility
Site de production
Produktionsstätte
Stabilimento di produzione
Oficina de producción
Productiefaciliteit
Производственное подразделение
OMRON (DALIAN) CO., LTD.
Economic & Technical Development Zone
Dalian 116600, CHINA
OMRON HEALTHCARE UK LIMITED
Opal Drive
Fox Milne Milton Keynes, MK15 0DG U.K.
Subsidiary
Succursale
Niederlassung
Consociata
Empresa filial
Dochteronderneming
Дочерняя компания
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
John-Deere-Str. 81a 68163 Mannheim GERMANY
www.omron-medizintechnik.de
OMRON Santé France SAS
14, rue de Lisbonne 93561 Rosny-sous-Bois Cedex, FRANCE
Prodotto in Cina
Сделано в КитаеHergestellt in China
Geproduceerd in ChinaFabriqué en Chine
Fabricado en China
Made in China
Русский
1

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Inhalt der Seiten


  • Page 1

    HEM-7200-E_B_M01_100917.pdf

    Check following components!
    Manufacturer

    OMRON HEALTHCARE Co., Ltd.
    24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto
    615-0084 JAPAN

    EU-representative

    OMRON HEALTHCARE EUROPE B.V.
    Kruisweg 577, 2132 NA Hoofddorp THE NETHERLANDS
    www.omron-healthcare.com

    Vérifier les composants suivants !
    Controleer de volgende onderdelen!
    Prüfen Sie folgende Teile des Lieferumfangs! Проверьте следующие компоненты!
    Controllare i componenti indicati di seguito!
    ¡Compruebe los siguientes componentes!

    English
    Français

    Production facility
    Oficina de producción
    OMRON (DALIAN) CO., LTD.
    Economic & Technical Development Zone
    Site de production
    Productiefaciliteit
    Produktionsstätte
    Производственное подразделение Dalian 116600, CHINA
    Stabilimento di produzione

    Subsidiary
    Succursale
    Niederlassung
    Consociata

    Empresa filial
    Dochteronderneming
    Дочерняя компания

    OMRON HEALTHCARE UK LIMITED
    Opal Drive
    Fox Milne Milton Keynes, MK15 0DG U.K.
    OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
    John-Deere-Str. 81a 68163 Mannheim GERMANY
    www.omron-medizintechnik.de
    OMRON Santé France SAS
    14, rue de Lisbonne 93561 Rosny-sous-Bois Cedex, FRANCE
    Made in China
    Fabricado en China
    Fabriqué en Chine Geproduceerd in China
    Hergestellt in China
    Сделано в Китае
    Prodotto in Cina

    Deutsch
    Italiano

    Digital Automatic Blood Pressure Monitor

    Model M3
    Instruction Manual

    Español
    Nederlands
    Русский

    IM-HEM-7200-E-05-11/2010
    5329491-4C



  • Page 2

    Contents
    1

    2

    Thank you for purchasing the OMRON M3 Digital Automatic Blood Pressure Monitor.
    The OMRON M3 is a compact, fully automatic blood pressure monitor, operating on the
    oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For
    comfortable controlled inflation without the need of pressure pre-setting or re-inflation the
    devices uses its advanced “IntelliSense” technology.
    Intended Use
    This product is designed to measure the blood pressure and pulse rate of people within the
    range of the designated arm cuff, following the instructions in this instruction manual.
    It is mainly designed for general household use. Please read the Important Safety Information
    in this instruction manual before using the unit.

    b

    F
    J
    L
    c

    G

    A

    B
    F

    H

    C

    E

    D
    d

    a

    Important Safety Information ........................1
    1. Overview ...................................................3
    2. Preparation ...............................................4
    2.1 Installing/Replacing the Batteries...........4
    2.2 Setting the Buzzer/ Date and Time ........5
    3. Using the Unit ...........................................6
    3.1 Applying the Arm Cuff ............................6
    3.2 How to Sit Correctly ...............................7
    3.3 Taking a Reading ...................................8
    3.4 Using the Memory Function .................10

    4. Troubleshooting and Maintenance .......12
    4.1 The Icons and Error Messages ........... 12
    4.2 Troubleshooting................................... 14
    4.3 Maintenance ........................................ 16
    4.4 Storage ................................................ 17
    5. Optional Parts .........................................18
    6. Technical Data ........................................19
    7. Some Useful Information about Blood Pressure.... 22

    Please read this instruction manual thoroughly before using the unit.
    Please keep for future reference.
    For specific information about your own blood pressure, CONSULT YOUR
    DOCTOR.

    I

    J

    K

    M

    U

    N

    V

    O
    P
    L

    Q
    R
    S
    T

    W
    X
    Y
    Z



  • Page 3

    HEM-7200-E_A_M_main.book Page 1 Wednesday, September 29, 2010 10:27 AM

    Important Safety Information
    Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis.
    Please read this section carefully before using the unit.
    Warning:
    • Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
    (General Usage)
    • Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.
    • People with severe blood flow problems, or blood disorders, should consult a doctor before using the unit, as cuff
    inflation can cause internal bleeding.
    (AC Adapter Usage)
    • Never plug in or unplug the power cord from the electric outlet with wet hands.
    (Battery Usage)

    EN

    • If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult a doctor immediately.
    Caution:
    • Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user
    or patient or damage to the equipment or other property.
    (General Usage)
    • Do not leave the unit unattended with infants or persons who cannot express their consent.
    • Do not use the unit for any purpose other than measuring blood pressure.
    • Do not disassemble the unit or arm cuff.
    • Use only the approved arm cuff for this unit. Use of other arm cuffs may result in incorrect measurement results.
    • Make sure that the air tube is not wrapped around other parts of your body when taking measurements at night.
    This could result in injury when the air pressure in the air tube is increased.
    • Do not leave the cuff wrapped on the arm if taking measurements during the night. This could result in injury.
    • Do not inflate the arm cuff over 299 mmHg.
    • Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in
    incorrect operation of the unit.
    • Do not operate the unit in a moving vehicle (car, airplane).
    • To inflate the cuff manually, refer to Section 3.3. If the cuff is over inflated, it can cause internal bleeding.

    1



  • Page 4

    HEM-7200-E_A_M_main.book Page 2 Wednesday, September 29, 2010 10:27 AM

    Important Safety Information
    (AC Adapter Usage)
    • Use only the original AC adapter (optional) designed for this unit. Use of unsupported adapters may damage and/or
    may be hazardous to the unit.
    • Plug the AC adapter into the appropriate voltage outlet. Do not use a multiple-tap.
    • Do not use the AC adapter if the unit or the power cord is damaged. Turn off the power and unplug the power cord
    immediately.
    (Battery Usage)







    If battery fluid should get on your skin or clothing, immediately rinse with plenty of clean water.
    Use only four “AA” alkaline or manganese batteries with this unit. Do not use other types of batteries.
    Do not insert the batteries with their polarities incorrectly aligned.
    Replace old batteries with new ones immediately. Replace all four batteries at the same time.
    Remove the batteries if the unit will not be used for three months or more.
    Do not use new and used batteries together.

    General Precautions
    • Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
    • Do not take measurements after bathing, drinking alcohol, smoking, exercising or eating.
    • Do not forcibly bend the arm cuff or bend the air tube excessively.
    • When removing the air tube, pull on the air plug at the connection with the main unit not the tube itself.
    • Do not inflate the arm cuff when it is not wrapped around your arm.
    • Do not wash the arm cuff or immerse it in water.
    • Read and follow the “Important information regarding Electro Magnetic Compatibility (EMC)” in the Technical Data
    Section.
    • Read and follow the “Correct Disposal of This Product” in the Technical Data Section when disposing of the device
    and any used accessories or optional parts.

    2



  • Page 5

    HEM-7200-E_A_M_main.book Page 3 Wednesday, September 29, 2010 10:27 AM

    1. Overview
    Open the rear cover page to read following:

    The alphabet in the rear cover page correspond to those in the body page.

    Main unit
    Display
    O/I START ( ) button
    Memory button
    Up/Down (
    ) buttons
    Buzzer and Date/Time setting
    (
    ) button
    F Air Jack
    G Battery compartment
    H AC adapter jack (for optional AC
    adapter)
    A
    B
    C
    D
    E

    Arm cuff

    Arm cuff (Medium cuff: arm
    circumference 22-32cm)
    J Coloured Marker
    K Air tube
    L Air plug
    I

    Display
    M Systolic blood pressure
    N Diastolic blood pressure
    O Heartbeat symbol
    (Flashes during measurement)

    EN

    P Memory symbol
    Q Deflation symbol
    R Average value symbol
    S Movement error symbol
    T Date/Time display
    U Blood pressure level indicator
    V Pulse display
    W Buzzer symbol
    X

    Y
    Z

    Cuff wrapping guide
    Battery low symbol
    Irregular heartbeat symbol

    3



  • Page 6

    HEM-7200-E_A_M_main.book Page 4 Wednesday, September 29, 2010 10:27 AM

    2. Preparation
    2.1 Installing/Replacing the
    Batteries
    1. Remove the battery cover.
    2. Insert four “AA” batteries as
    indicated in the battery
    compartment and then replace the battery cover.
    Notes:
    • If the battery low symbol (
    ) appears on the display, turn the unit off then replace all batteries at the same
    time.
    • The measurement values continue to be stored in memory even after the batteries are replaced.

    Disposal of used batteries should be carried out in accordance with the national
    regulations for the disposal of batteries.

    4



  • Page 7

    HEM-7200-E_A_M_main.book Page 5 Wednesday, September 29, 2010 10:27 AM

    2. Preparation

    2.2 Setting the Buzzer/ Date and Time
    1. Press the
    button.
    2. Set the Buzzer.
    1)
    2)

    Push
    or
    buttons to
    select On or Off.
    Push the

    EN

    button to confirm.

    Off

    On

    3. Set the monitor to the correct date and time before
    taking a measurement for the first time.
    1)

    Push
    or
    change.

    2)

    Push the

    buttons to
    button to confirm.

    Next setting appears.

    Year

    Month

    Day

    Hour

    Minute

    4. Press the O/I START button to store the setting.
    Notes:

    • If the batteries have been removed for 30 seconds or more, the date and time setting will need to be reset.
    • If the date and time are not set, “-:--” appears during or after measurement.

    5



  • Page 8

    HEM-7200-E_A_M_main.book Page 6 Wednesday, September 29, 2010 10:27 AM

    3. Using the Unit
    Open both the front and rear covers to read following:
    The alphabet and number in the cover page correspond to those in the body page.

    3.1 Applying the Arm Cuff
    Remove tight-fitting clothing or tight rolled up sleeve from your upper arm.
    Do not place the cuff over thick clothes.

    1. Insert L the air plug into F the air jack.
    2. Put your arm through the cuff loop.
    3. Position the arm correctly.
    The bottom edge of the cuff should be a 1 to 2 cm above the elbow. J Marker (arrow
    under tube) is centred on the middle of your inner arm. Close the fabric fastener FIRMLY.
    Notes:
    • When you take a measurement on the right arm, air tube will be at the side of your elbow. Be careful not
    to rest your arm on the air tube. --- b
    • The blood pressure can differ between the right arm and the left arm, and therefore also the measured
    blood pressure values can be different. Omron recommends to always use the same arm for
    measurement. If the values between the two arms differ substantially, please check with your physician
    which arm to use for your measurement.

    6



  • Page 9

    HEM-7200-E_A_M_main.book Page 7 Wednesday, September 29, 2010 10:27 AM

    3. Using the Unit

    3.2 How to Sit Correctly
    To take a measurement, you need to be relaxed and comfortably seated, under
    comfortable room temperature. No eating, smoking or exercising 30 minutes before
    taking a measurement.
    • Sit on a chair with your feet flat on the floor.
    • Sit upright with your back straight. --- c
    • The cuff should be at the same level as your heart. --- d

    Cuff wrapping guide

    EN

    The Cuff Wrapping Guide is a unique feature that indicates if the cuff is not wrapped tightly enough around
    the arm. Even when the
    is displayed, a blood pressure reading will be taken.

    Note:
    This reading is NOT reliable due to the incorrect wrapping of the cuff. Please wrap the cuff again, taking
    care to wrap it correctly and take the measurement again.
    When the
    is displayed, the cuff is correctly wrapped tightly enough on the arm and the reading is
    accurate and reliable.

    7



  • Page 10

    HEM-7200-E_A_M_main.book Page 8 Wednesday, September 29, 2010 10:27 AM

    3. Using the Unit

    3.3 Taking a Reading
    Notes:

    • To cancel a measurement, press the O/I START button to turn off the unit and to release the air in the arm cuff.
    • Remain still while taking a measurement.

    1. Press the O/I START button.
    The cuff will start to inflate automatically.
    The buzzer will sound when measurement is completed.
    START

    INFLATE

    DEFLATE

    END

    COMPLETE

    If your systolic pressure is
    more than 220 mmHg
    After the cuff starts to inflate, press and hold
    the O/I START button until the monitor
    inflates 30 to 40 mmHg higher than your
    expected systolic pressure.

    Notes:
    • The monitor will not inflate above 300 mmHg.
    • Do not apply more pressure than necessary.

    Notes:
    • During measurement, the buzzer (if set to “on”) will beep in rhythm with your heartbeat.
    • Wait 2-3 minutes before taking another blood pressure measurement. Waiting between readings allows the
    arteries to return to the condition prior to taking the blood pressure measurement.
    Warning:
    Self-diagnosis of measured results and treatment are dangerous. Please follow the instructions of
    your doctor.

    8



  • Page 11

    HEM-7200-E_A_M_main.book Page 9 Wednesday, September 29, 2010 10:27 AM

    3. Using the Unit

    2. Remove the arm cuff.
    3. Press the O/I START button to turn off the monitor.
    The monitor automatically stores the measurement in its memory.
    It will automatically turn off after five minutes.
    Important:
    • Recent research suggests that the following values can be used as a guide to
    high blood pressure for measurements taken at home.
    Systolic Blood Pressure

    Above 135 mmHg

    Diastolic Blood Pressure

    Above 85 mmHg

    EN

    This criteria is for home blood pressure measurement.
    For professional office blood pressure measurement criteria, please refer to
    Chapter 7.
    • Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats
    can influence the results of the measurement. The irregular heartbeat algorithm
    automatically determines if the measurement is usable or needs to be repeated. If the
    measurement results are affected by irregular heartbeats but the result is valid, the result is
    shown together with the irregular heartbeat symbol (
    ). If the irregular heartbeats cause
    the measurement to be invalid, no result is shown. If the irregular heartbeat symbol (
    )
    is shown after you have taken a measurement, repeat the measurement. If the irregular
    heartbeat symbol (
    ) is shown frequently, please make your doctor aware of it.

    9



  • Page 12

    HEM-7200-E_A_M_main.book Page 10 Wednesday, September 29, 2010 10:27 AM

    3. Using the Unit

    3.4 Using the Memory Function
    The monitor automatically stores the result up to 60 sets. It can also calculate an
    average reading based on the measurements from the last three readings taken
    within 10 minutes. If there are only two readings in memory for that period, the
    average will be based on two readings. If there is one reading in memory for that
    period, the average will be based on one reading.
    Notes:

    • If the memory is full, the monitor will delete the oldest readings.
    • When viewing the reading taken without setting the date and time, “-:--” is displayed instead of the date
    and time.
    Average Value
    Previous Reading
    Alternating date/time display

    Old

    10

    View the Average
    Value

    View the Previous
    Reading

    • If the previous reading
    was taken without
    setting the date and
    time, the average
    value is not calculated.

    • The Memory number
    appears for a second
    before the pulse rate is
    displayed. The newest
    set is numbered “1”.

    Recent

    View the Readings
    Stored in Memory



  • Page 13

    HEM-7200-E_A_M_main.book Page 11 Wednesday, September 29, 2010 10:27 AM

    3. Using the Unit
    To delete all the values stored in memory
    When the memory symbol ( ) appears, first press the Memory button. Then while
    holding it down, press the O/I START button simultaneously for about 2-3 seconds.

    EN
    Note:
    You cannot partially delete the stored readings.

    11



  • Page 14

    HEM-7200-E_A_M_main.book Page 12 Wednesday, September 29, 2010 10:27 AM

    4. Troubleshooting and Maintenance
    4.1 The Icons and Error Messages
    Error Display

    12

    Cause

    Remedy

    Irregular or weak pulses are detected.

    Remove the arm cuff. Wait 2-3 minutes and then
    take another measurement.
    Repeat the steps in section 3.3. If this error
    continues to appear, contact your doctor.

    Movement during measurement.

    Carefully read and repeat the steps in section 3.3.

    Cuff is not applied correctly.

    Apply the arm cuff correctly. Refer to section 3.1.

    Blink

    The batteries are low.

    You should replace them with new ones ahead of
    time.
    Refer to section 2.1.

    Lit

    The batteries are exhausted.

    You should replace them with new ones at once.
    Refer to section 2.1.



  • Page 15

    HEM-7200-E_A_M_main.book Page 13 Wednesday, September 29, 2010 10:27 AM

    4. Troubleshooting and Maintenance
    Error Display

    Cause

    Remedy

    Cuff is under inflated.

    Carefully read and repeat the steps listed under
    section 3.3.

    Movement during measurement.

    Repeat measurement. Remain still and do not
    talk during measurement.
    Refer to section 3.3.

    Air plug disconnected.

    Insert the air plug securely.
    Refer to section 3.1.

    Arm cuff not applied correctly.

    Apply the arm cuff correctly.
    Refer to section 3.1.

    Clothing is interfering with the arm cuff.

    Remove any clothing interfering with the arm cuff.
    Refer to section 3.1.

    Air is leaking from the arm cuff.

    Replace cuff with new one.
    Refer to Chapter 5.

    The arm cuff was inflated above 299
    mmHg when inflating the cuff manually.

    Do not inflate the arm cuff above 299 mmHg.
    Refer to section 3.3.

    Device error.

    Contact your OMRON retail outlet or distributor.

    EN

    Note:
    The irregular heartbeat symbol (

    ) may also be displayed with error messages.

    13



  • Page 16

    HEM-7200-E_A_M_main.book Page 14 Wednesday, September 29, 2010 10:27 AM

    4. Troubleshooting and Maintenance

    4.2 Troubleshooting
    Problem

    Cause
    Arm cuff not applied correctly.

    Apply the arm cuff correctly.
    Refer to section 3.1.

    Movement or talking during
    measurement.

    Remain still and do not talk during
    measurement.
    Refer to section 3.3.

    Clothing is interfering with the arm
    cuff.

    Remove any clothing interfering
    with the arm cuff.
    Refer to section 3.2.

    The air tube is not securely
    connected into the main unit.

    Make sure that the air tube is
    connected securely.
    Refer to section 3.1.

    Air is leaking from the arm cuff.

    Replace the arm cuff with a new
    one.
    Refer to Chapter 5.

    Arm cuff deflates too soon.

    The arm cuff is loose.

    Apply the cuff correctly so that it is
    firmly wrapped around the arm.
    Refer to section 3.1.

    Cannot measure or readings are too
    low or too high.

    The arm cuff has not been inflated
    sufficiently.

    Inflate the cuff so that it is 30 to
    40 mmHg above your previous
    measurement result.
    Refer to section 3.3.

    The reading is extremely low (or
    high).

    Arm cuff pressure does not rise.

    14

    Remedy



  • Page 17

    HEM-7200-E_A_M_main.book Page 15 Wednesday, September 29, 2010 10:27 AM

    4. Troubleshooting and Maintenance
    Problem

    Nothing happens when you press the
    buttons.

    Other problems.

    Cause

    Remedy

    The batteries are empty.

    Replace the batteries with new
    ones.
    Refer to section 2.1.

    The batteries have been inserted
    incorrectly.

    Insert the batteries with the correct
    (+/-) polarity.
    Refer to section 2.1.

    • Press the O/I START button and repeat measurement.
    • If the problem continues, try replacing the batteries with new ones.

    EN

    If this still does not solve the problem, contact your OMRON retail outlet or
    distributor.

    15



  • Page 18

    HEM-7200-E_A_M_main.book Page 16 Wednesday, September 29, 2010 10:27 AM

    4. Troubleshooting and Maintenance

    4.3 Maintenance
    To protect your unit from damage, please observe the following:










    Do not subject the main unit and the cuff to extreme temperatures, humidity, moisture or direct sunlight.
    Do not fold the cuff or tubing tightly.
    Do not inflate the arm cuff over 299 mmHg.
    Do not disassemble the unit.
    Do not subject the unit to strong shocks or vibrations (for example, dropping the unit on the floor).
    Do not use volatile liquids to clean the main unit.
    Do not wash the arm cuff or immerse it in water.
    Do not use petrol, thinners or similar solvents to clean the arm cuff.
    Do not carry out repairs of any kind yourself. If a defect occurs, consult your OMRON retail outlet or
    distributor as mentioned on the packaging.

    • The unit should be cleaned with a soft, dry cloth.
    • Use a soft, moistened cloth and soap to clean the arm cuff.

    Calibration and Service
    • The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service
    life.
    • It is generally recommended to have the unit inspected every two years to ensure correct functioning and
    accuracy. Please consult your authorised OMRON dealer or the OMRON Customer Service at the
    address given on the packaging or attached literature.

    16



  • Page 19

    HEM-7200-E_A_M_main.book Page 17 Wednesday, September 29, 2010 10:27 AM

    4. Troubleshooting and Maintenance

    4.4 Storage
    Keep the unit in its storage case when not in use.

    1. Unplug the air tube from the air connector.
    2. Gently fold the air tube into the arm cuff.
    Note:

    EN

    Do not bend the air tube excessively.

    3. Place the arm cuff and main unit in the
    storage case.
    Do not store the unit in the following situations:
    • If the unit is wet.
    • Locations exposed to extreme temperatures, humidity, direct sunlight, dust or
    corrosive vapours.
    • Locations exposed to vibrations, shocks or where it will be at an angle.

    17



  • Page 20

    HEM-7200-E_A_M_main.book Page 18 Wednesday, September 29, 2010 10:27 AM

    5. Optional Parts
    Medium Arm Cuff

    Large Arm Cuff

    Arm circumference 22 - 32 cm

    Arm circumference 32 - 42 cm

    CM2-9513256-6 (Model: HEM-CR24)
    CM1-9997578-9

    CL2-9513255-8 (Model: HEM-CL24)
    CL1-9996760-3

    AC Adapter S

    Adapter S-9515336-9

    Using the Optional AC Adapter

    1. Insert the AC adapter plug
    into the AC adapter jack on
    the rear side of the main unit.

    2. Plug the AC adapter into an
    electrical outlet.
    To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first
    and then remove the AC adapter plug from the main unit.

    18



  • Page 21

    HEM-7200-E_A_M_main.book Page 19 Wednesday, September 29, 2010 10:27 AM

    6. Technical Data
    Product Description
    Model
    Display
    Measurement Method
    Measurement Range
    Accuracy
    Inflation
    Deflation
    Memory
    Rating
    Power Source
    Battery Life
    Applied Part

    Digital Automatic Blood Pressure Monitor
    OMRON M3 (HEM-7200-E/ HEM-7200-E8)
    LCD Digital Display
    Oscillometric method
    Pressure: 0 mmHg to 299 mmHg
    Pulse: 40 to 180/min.
    Pressure: ±3 mmHg
    Pulse: ± 5% of display reading
    Fuzzy-logic controlled by electric pump
    Automatic pressure release valve
    60 Measurements with date and time
    DC6V
    4W
    4 “AA” batteries 1.5V or AC/ DC adapter
    (Adapter S-9515336-9 (optional), INPUT AC100-240V 50/60Hz 0.12A)
    Capacity of new alkaline batteries is approx. 1500 measurements

    EN

    = Type
    TypeBB
    =

    Protection Against Electric
    Shock

    Internally powered ME equipment (When using only the batteries)

    Operating temperature/ Humidity
    Storage temperature/ Humidity/
    Air pressure
    Console Weight
    Cuff Weight
    Outer Dimensions
    Cuff Dimensions
    Cuff Material
    Package Content

    +10°C to +40°C / Maximum: 30 to 90% RH

    = Class II ME equipment (Optional AC adapter)

    -20°C to +60°C / Maximum: 10 to 95% RH / 700-1060hPa
    Approximately 340g without batteries
    Approximately 120g
    Approximately 123 (w) mm × 85 (h) mm × 141(l) mm
    Approximately 146 mm × 446 mm (Medium cuff: arm circumference 22 to 32 cm)
    Nylon and polyester
    Main unit, cuff, instruction manual, storage case, battery set, guarantee card, blood
    pressure pass

    Note: Subject to technical modification without prior notice.

    19



  • Page 22

    HEM-7200-E_A_M_main.book Page 20 Wednesday, September 29, 2010 10:27 AM

    6. Technical Data

    • This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive).
    • This blood pressure monitor is designed according to the European Standard EN1060, Noninvasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary
    requirements for electromechanical blood pressure measuring systems.
    • This OMRON product is produced under the strict quality system of OMRON HEALTHCARE
    Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the
    Pressure Sensor, is produced in Japan.

    20



  • Page 23

    HEM-7200-E_A_M_main.book Page 21 Wednesday, September 29, 2010 10:27 AM

    6. Technical Data
    Important information regarding Electro Magnetic Compatibility (EMC)
    With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in
    use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result
    in incorrect operation of the medical device and create a potentially unsafe situation.
    Medical devices should also not interfere with other devices.
    In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product
    situations, the EN60601-1-2:2007 standard has been implemented. This standard defines the levels of immunity to
    electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.
    This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for
    both immunity and emissions.
    Nevertheless, special precautions need to be observed:
    • Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic
    fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe
    situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case
    the distance is shorter.
    Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at
    the address mentioned in this instruction manual.
    Documentation is also available at www.omron-healthcare.com.

    EN

    Correct Disposal of This Product
    (Waste Electrical & Electronic Equipment)
    This marking shown on the product or its literature, indicates that it should not be disposed of, with other
    household wastes at the end of its working life. To prevent possible harm to the environment or human
    health from uncontrolled waste disposal, please separate this from other types of wastes and recycle it
    responsibly to promote the sustainable reuse of material resources.
    Household users should contact either the retailer where they purchased this product, or their local government
    office, for details of where and how they can take this item for environmentally safe recycling.
    Business users should contact their supplier and check the terms and conditions of the purchase contract. This
    product should not be mixed with other commercial wastes for disposal.
    This product does not contain any hazardous substances.

    21



  • Page 24

    HEM-7200-E_A_M_main.book Page 22 Wednesday, September 29, 2010 10:27 AM

    7. Some Useful Information about Blood Pressure
    What is Blood Pressure?
    Blood pressure is a measure of the force of blood flowing against the walls of the arteries.
    Arterial blood pressure is constantly changing during the course of the heart’s cycle.
    The highest pressure in the cycle is called the Systolic Blood Pressure; the lowest is the
    Diastolic Blood Pressure.
    Both pressure readings, the Systolic and Diastolic, are necessary to enable a doctor to
    evaluate the status of a patient’s blood pressure.

    What is Irregular Heartbeat?
    An irregular heartbeat is a heartbeat rhythm that varies by
    more than 25% from the average heartbeat rhythm detected
    while the unit is measuring the systolic and diastolic blood
    pressure.
    If such an irregular rhythm is detected more than twice
    during measurement, the irregular heartbeat symbol (
    )
    appears on the symbol when the measurement results are
    displayed.

    Normal Heartbeat
    Pulse
    Blood pressure

    Irregular Heartbeat
    Short

    Long

    Pulse
    Blood pressure

    What is Arrhythmia?
    Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in the bioelectrical system that drives the heartbeat. Typical symptoms are skipped heartbeats,
    premature contraction, an abnormally rapid (tachycardia) or slow (bradycardia) pulse.

    22



  • Page 25

    HEM-7200-E_A_M_main.book Page 23 Wednesday, September 29, 2010 10:27 AM

    7. Some Useful Information about Blood Pressure
    Why is it a Good Thing to measure Blood Pressure at Home?
    Upper curve: systolic blood pressure
    Lower curve: diastolic blood pressure

    mmHg

    Many factors such as physical activity,
    200
    anxiety, or the time of day, can
    influence your blood pressure. A single
    150
    measurement may not be sufficient for
    an accurate diagnosis. Thus it is best to
    100
    try and measure your blood pressure at
    the same time each day, to get an
    50
    accurate indication of any changes in
    blood pressure. Blood pressure is
    typically low in the morning and
    increases from afternoon to evening. It
    is lower in the summer and higher in the winter.

    EN
    6

    12

    18

    24

    Time of day
    Example: fluctuation within a day (male, 35 years old)

    23



  • Page 26

    HEM-7200-E_A_M_main.book Page 24 Wednesday, September 29, 2010 10:27 AM

    7. Some Useful Information about Blood Pressure
    Classification of Blood Pressure by the World Health Organization
    The World Health Organization (WHO) and the
    International Society of Hypertension (ISH) developed
    the Blood Pressure Classification shown in this figure.
    This classification is based on the blood pressure
    values measured on people in a sitting position in
    outpatient departments of hospitals.
    Note:
    There is no universally accepted definition of hypotension. However,
    those having the systolic pressure below 100 mmHg are assumed
    as hypotensive.

    24

    Systolic blood pressure
    (mmHg)

    Severe hypertension
    180
    Moderate hypertension
    160
    Mild hypertension
    140
    Normal systolic value
    130
    120

    Normal blood
    pressure
    Optimal blood
    pressure
    (target value)
    80

    85

    90

    100 110

    Diastolic
    blood
    pressure
    (mmHg)






Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Omron M3 Intellisense - HEM-7051-E wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Holen Sie sich Ihr Benutzerhandbuch per E-Mail

Geben Sie Ihre E-Mail-Adresse ein, um das Handbuch zu erhalten von Omron M3 Intellisense - HEM-7051-E in der Sprache / Sprachen: Englisch als Anhang in Ihrer E-Mail.

Das Handbuch ist 1,45 mb groß.

 

Sie erhalten das Handbuch in Ihrer E-Mail innerhalb von Minuten. Wenn Sie keine E-Mail erhalten haben, haben Sie wahrscheinlich die falsche E-Mail-Adresse eingegeben oder Ihre Mailbox ist zu voll. Darüber hinaus kann es sein, dass Ihr ISP eine maximale Größe für E-Mails empfangen kann.

Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info