ERROR MESSAGES AND DETAILS ABOUT YOUR SYSTEM
Detailed information about your system
Comparing meter and laboratory results
Test results with the OneTouch® UltraMini® Meter are plasma-calibrated. This helps you and your healthcare professional to compare your
meter results with laboratory tests. If you have been using another type of meter—one that provides whole blood-calibrated results—you
may notice that your test results with the OneTouch® UltraMini® Meter are approximately 12% higher. OneTouch® UltraMini® Meter test results
and laboratory test results both are expressed in plasma-equivalent units. However, your meter result may dier from your laboratory result
due to normal variation. Meter results can be aected by factors and conditions that do not aect laboratory results in the same way.
Your OneTouch® UltraMini® Meter glucose value is considered accurate when it is within ±20% of the laboratory measurement. There are some
specic situations that could cause a dierence of more than ±20%:
You have eaten recently. The blood glucose level from a ngertip can be up to 70 mg/dL higher than blood drawn from a vein
(venous sample) used for a lab test.
1
• Your hematocrit (percentage of blood that is red blood cells) is high (above 55%) or low (below 30%).
• You are suering from severe dehydration.
You tested at a temperature near the low end of the operating range (43°F) and you get a high glucose result (i.e., greater than
180 mg/dL). In this situation, repeat the test in a warmer environment with a new test strip as soon as possible.
For accuracy and precision data and for important information on limitations, see the insert that comes with your test strips. To maximize your
chances of an accurate comparison between meter and laboratory results, follow a few basic guidelines:
Before going to the lab
Perform a control solution test to make sure the meter is
working properly.
Do Not eat for at least eight hours before you test your blood.
• Take your meter with you to the lab.
While at the lab
• Conduct your meter test within 15 minutes of the lab test.
• Use only fresh, capillary blood obtained from the ngertip.
Follow all instructions in this User Guide for performing a
blood glucose test with your meter.
1. Sacks, D.B.: “Carbohydrates.” Burtis, C.A., and Ashwood, E.R. (ed.), Tietz Textbook of Clinical Chemistry. Philadelphia: W.B. Saunders Company (1994), 959.
Technical specications
Reported result range 20–600 mg/dL
Calibration Plasma-equivalent
Sample Fresh capillary whole blood
Test time 5 seconds
Assay method Glucose oxidase biosensor
Meter power source One replaceable 3.0 Volt CR 2032
lithium battery (or equivalent)
Unit of measure mg/dL
Memory 500 blood glucose test results
Automatic shuto 2 minutes after last action
Size 4.25 x 1.26 x .67 inches
Weight Approximately 1.4 ounces, with battery
Operating ranges
Temperature: 43–111°F
Altitude: up to 10,000 feet
Relative humidity: 10–90%
Hematocrit: 30–55%
Battery ratings One 3.0 V d.c., 3 mA
(one CR 2032 battery)
direct current
Symbols
Cautions and Warnings: Refer to safety-related notes in the User Guide and inserts that came with your meter and testing supplies.
Direct current Low battery
SN
Serial Number
Electrical and safety standards
The meter has been tested for immunity to Level 4 electrostatic discharge as specied in IEC 61000-4-2. This meter has been tested for
immunity to radio frequency interference over the frequency range 80 MHz to 2.5 GHz at 3 V/m as specied in IEC 61000-4-3. Degree of
protection rating: Minimum of IP2X. This meter complies with CISPR 11:2003, Class B (Radiated Only). Emissions of the energy used are low and
are not likely to cause interference in nearby electronic equipment.
C US
CAN/CSA C22.2 61010-1:04, UL 61010-1:04, IEC 61010-1 and IEC 61010-2-101.
Guarantee
LifeScan guarantees that the OneTouch® UltraMini® Meter will be free of defects in material and workmanship for three years, valid from the
date of purchase. The guarantee extends only to the original purchaser and is not transferable.
27

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für OneTouch Ultra Mini wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von OneTouch Ultra Mini

OneTouch Ultra Mini Bedienungsanleitung - Französisch - 31 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info