Rossmax Neb Tester is a portable and easy-to-use device that allows the user to quickly
check the performance of the pump of a compressor nebulizer. Power source is not
required for Neb Tester during operation and the test result is easy to read and interpret.
Neb Tester consists of an oil pressure gauge, ow meter, air tube and stainless steel stand.
This device is used to simulate the working conditions of the pump when the nebulizer is
in operation with a medication cup, and you are able to check:
1. Maximum (free) air ow
2. Operational air ow at a particular pressure
3. Maximum air pressure
Parts name/location
Introduction
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
Flow meter
Regulator
Pressure
gauge
Air tube
Insert air tube into connectorPlace the device in a
horizontal position
Open the rubber vent-plug
and then close it after
Back view
Installation
Operation procedures
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
1. Check maximum (free) air ow
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Adjust air ow and set the level of the pressure at 0 PSI by turning the
regulator of ow meter.
Step 6: Check the air ow rate.
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
2. Check operational air ow at a particular pressure
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Set the pressure to a desired level by adjusting the regulator of the ow
meter.
Step 6: Check the air ow rate.
Pressure (PSI) Flow rate (L/min)
NA100 / NB500 (N1) 15.0 ≥ 3.5
NE100 / NF100/ NJ100 (N4) 15.0 ≥ 3.5
NK1000/ NB80 / NF80 (N1) 8.0 ≥ 3.0
NB60 / NI60/ NH60 / NF60 (N1) 8.0 ≥ 2.0
NL100 (N2) 10.5 ≥ 3.5
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
3. Check maximum air pressure
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Set the air ow rate at 0 L/min by turning the regulator of ow meter
clockwise completely.
Step 6: Check the values on the pressure gauge.
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
Air
compressor
connector
Rubber
vent-plug
2

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Rossmax Neb Tester wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info