712111
1
Zoom out
Zoom in
Vorherige Seite
1/2
Nächste Seite
Pairing the Blood Pressure Monitor with your Smart-
phone
To begin using Bluetooth® for the rst time, please visit the
website at http://www.rossmax.com for the initial set-up
instructions.
1. Download and install the free APP onto your smart-
phone.
2. To pair this device with your smartphone, turn on the
device, Bluetooth® and the App of smartphone, and fol-
low set-up and pairing instructions.
3. The Bluetooth® symbol ( ) will ash on the display that
indicates the pairing is proceeding.
- If the pairing is successful, the Bluetooth® symbol ( ) will stop ashing and appear on
the display, the device will start the measurement automatically. The current measured
reading (Systolic, Diastolic and Pulse) will automatically be transferred to the App.
- If the pairing has failed, the Bluetooth® symbol ( ) will continue ashing until the device
shuts o automatically after 1 minute. Please re-pair this device with your smartphone.
Notes: 1. You can pair up to 2 destination devices with the monitor
2. Unpairing your device will not delete the information on the phone App.
3. If you re-pair your smartphone with your blood pressure monitor, all prior reading history
stored on the phone App will be retained.
4. Bluetooth® data transfer will reduce the battery capacity.
APP
APP
APP
APP
120mmHg
120mmHg
USB
PC
BP Manager
APP
APP
APP
APP
120mmHg
120mmHg
USB
PC
BP Manager
1. The monitor has two memory zones (1 and 2). Each zone can store up to 60 measure-
ments.
2. To read memory values from a selected memory zone, use the User-Switching key to
select a memory zone (1 or 2) from which you want to recall values. Press the Memory
key. The rst reading displayed is the average of all morning readings from the last 7
days.
3. Continue to press the Memory key to view the average of all nighttime readings from
the last 7 days.
4. Press the Memory key again to view the average of the last 3 measurements stored in
memory, and the last previously stored measurement. Every measurement comes with
an assigned memory sequence number.
Note: The memory bank can store up to 60 readings per memory zone. When the num-
ber of readings exceeds 60, the oldest data will be replaced with the new record.
Note: AM is dened as 4:00 AM – 11:59 AM
Note: PM is dened as 6:00 PM – 2:00 AM
1. Press the User-Switching key to select memory zone 1 or memory zone 2.
2. Press and hold the Memory key for approximately 5 seconds, then the data in the mem-
ory zone can be erased automatically.
1. To adjust the date/time/Bluetooth® in the monitor after installing or replacing batter-
ies. The display will show a blinking number showing the year.
2. Change the year by pressing the Memory key, each press will increase the number.
Press the ON/OFF/START key to conrm the entry and the screen will show a blinking
number representing the date.
3. Change the date, the hour and the minute as described in Step 2 above, using the
Memory key to change and the ON/OFF/START key to conrm the entries.
4. After adjusting the date/time, the Bluetooth® symbol (
) and the blinking icon “
will be shown on the display simultaneously. Use the Memory key to choose whether
automatic Bluetooth® data transfer is activated (Bluetooth® symbol ( )+ ) or de-
activated (Bluetooth® symbol ( )+ ) and conrm with the ON/OFF/START key.
5. Press the ON/OFF/START key again, “0” will reappear as the Blood Pressure Monitor is
ready for measurement.
Rossmax provides a free, integrated and user-friendly blood pressure management soft-
ware which can be downloaded and installed on your computer. You may purchase a spe-
cial designed USB cable in order to connect Rossmax’s blood pressure monitor and your
PC. Please visit the website at http://www.rossmax.com for proceeding the downloading
and installation process.
If any abnormality will arise during use, please check the following points.
Symptoms Check Points Correction
No display when
the ON/OFF/START
key is pressed
Have the batteries run down? Replace them with four new batteries.
Have the batteries' polarities
been positioned incorrectly?
Re-insert the batteries in the correct
positions.
EE mark shown
on display or the
blood pressure
value is displayed
excessively low
(high)
Is the cuff placed correctly?
Wrap the cuff properly so that it is
positioned correctly.
Did you talk or move during
measurement?
Measure again. Keep wrist steady dur-
ing measurement.
Did you vigorously shake the
cuff during measurement?
Note: If the unit still does not work, return it to your dealer. Under no circumstance should
you disassemble and repair the unit by yourself.
1. The unit contains high-precision assemblies. Therefore, avoid extreme temperatures, hu-
midity, and direct sunlight. Avoid dropping or strongly shocking the main unit, and pro-
tect it from dust.
2. Clean the blood pressure monitor body and the cu carefully with a slightly damp, soft
cloth. Do not press. Do not wash the cu or use chemical cleaner on it. Never use thinner,
alcohol or petrol (gasoline) as cleaner.
3. Leaky batteries can damage the unit. Remove the batteries when the unit is not used for
a long time.
4. The unit should not be operated by children so to avoid hazardous situations.
5. If the unit is stored near freezing, allow it to acclimate at room temperature before use.
6. This unit is not eld serviceable. You should not use any tool to open the device nor should
you attempt to adjust anything inside the device. If you have any problems, please contact
the store or the doctor from whom you purchased this unit or please contact Rossmax
International Ltd.
7. As a common issue for all blood pressure monitors using the oscillometric measurement
function, the device may have diculty in determining the proper blood pressure for us-
ers diagnosed with diabetes, poor circulation of blood, kidney problems, or for users suf-
fered from stroke, or for unconscious users.
8. This unit is able to detect common arrhythmia (atrial or ventricular premature beats or
atrial brillation). The ARR, AFib and PC icons are displayed after the measurement if Atrial
Fibrillation and Premature Contraction was detected during the measurement. If ARR,
AFib or PC icons are displayed, you are advised to wait for a while and take another meas-
urement. It is strongly recommended that you consult your physician if the ARR, AFib or
PC icons appear often.
9. While the given device is able to detect specic pulse arrhythmia, the measurement ac-
curacy of the blood pressure meter may be impaired with the occurrence of pulse ar-
rhythmia.
10. To stop operation at any time, press the ON/OFF/START key, and the air in the cu will be
rapidly exhausted.
11. Once the ination reaches 300 mmHg, the unit will start deating rapidly for safety rea-
sons.
12. Please note that this unit can be a home healthcare product, but it is not intended to
serve as a substitute for the advice of a physician or medical professional.
13. Do not use this device for diagnosis or treatment of any health problem or disease. Meas-
urement results are for reference only. Consult a healthcare professional for interpreta-
tion of pressure measurements. Contact your physician if you have or suspect any medi-
cal problem. Do not change your medications without the advice of your physician or
healthcare professional.
Measurement Method Oscillometric
Measurement Range Pressure: 30~260 mmHg; Pulse: 40~199 beats/
minute
Pressure Sensor Semi conductor
Accuracy Pressure: ± 3 mmHg; Pulse: ± 5% of reading
Inflation Pump Driven
Deflation Automatic Air Release Valve
Memory capacity 60 memories for each zone x 2 zones
Auto-shut-off 1 minute after last key operation
Permissible Operating Temperature
and Humidity
10°C~40°C (50°F~104°F); 15%~85% RH; 700~1060
hPa
Permissible Transport and Storage
Temperature and Humidity
-10°C~60°C (14°F~140°F); 10%~90% RH; 700~1060
hPa
DC Power Source DC 6V four AAA Batteries
AC Power Source DC 6V, ≥600mA (Plug size: outer(-) is Ø4.0, in-
ner(+) is Ø1.7)
Dimensions 96 (L) X 139.7 (W) X 63.2 (H) mm
Weight 248.6g (G.W.) (w/o Batteries)
Arm circumference Adult: 24~40 cm (9.4”~15.7”)
Limited Users Adult users
:
Type BF: Device and cuff are designed to provide
special protection against electrical shocks.
IP Classification IP21: Protection against harmful ingress of water
and particulate matter
*Specifications are subject to change without notice.
Recalling Values from Memory
Clearing Values from Memory
Time and Bluetooth® Adjustment
Data Transfer via Bluetooth®
Data Transfer to PC
Troubleshooting
Cautionary Notes
Specications
2. Press the ON/OFF/START key. All digits will light up, checking the display functions. The
checking procedure will be completed in 2 seconds.
3. After all symbols appear, the display will show a blinking “0”. The monitor is ready to meas-
ure and will automatically inate the cu slowly to start measurement.
4. Bluetooth® will automatically be activated on the device, please see Data Transfer via
Bluetooth®. You can also start measurement directly by pressing the ON/OFF/START key
again, but the current measured reading (Systolic, Diastolic and Pulse) will not automati-
cally be transferred to the App.
5. When the measurement is completed, the cu will exhaust the pressure inside. Systolic
pressure, diastolic pressure and pulse will be shown simultaneously on the LCD screen.
The measurement is then automatically stored into the pre-designated memory zone.
6. In order to enhance the probability of pulse arrhythmia detection by the PARR technol-
ogy, measurement repetitions are recommended.
This monitor will re-inate automatically to approximately 220 mmHg if the system detects
that your body needs more pressure to measure your blood pressure.
Note: 1. This monitor automatically switches o approximately 1 minute after last key op-
eration.
2. To interrupt the measurement, simply press the ON/OFF/START key; the cu will deate
immediately.
3. During the measurement, do not talk or move your arm or hand muscles.
14. Electromagnetic interference: The device contains sensitive electronic components.
Avoid strong electrical or electromagnetic elds in the direct vicinity of the device (e.g.
mobile telephones, microwave ovens). These may lead to temporary impairment of
measurement accuracy.
15. Dispose of device, batteries, components and accessories according to local regulations.
16. This monitor may not meet its performance specication if stored or used outside tem-
perature and humidity ranges specied in Specications.
17. Please note that when inating, the functions of the limb in question may be impaired.
18. During the blood pressure measurement, blood circulation must not be stopped for an
unnecessarily long time. If the device malfunctions, remove the cu from the arm.
19. Avoid any mechanical restriction, compression or bending of the cu line.
20. Do not allow sustained pressure in the cu or frequent measurements. The resulting re-
striction of the blood ow may cause injury.
21. Ensure that the cu is not placed on an arm in which the arteries or veins are undergoing
medical treatment, e.g. intravascular access or therapy, or an arteriovenous (AV) shunt.
22. Do not apply the cu on the side, where a mastectomy has been performed in your
patient history.
23. Do not place the cu over wounds as this may cause further injury.
24. Only ever use the cus provided with the monitor or original replacement cus. Other-
wise erroneous results will be recorded.
25. Batteries can be fatal if swallowed. You should therefore store the batteries and products
where they are inaccessible to small children. If a battery has been swallowed, call a doc-
tor immediately.
26. Do not use the tubing and/or AC adaptor for any other purpose than those specied, as
they can cause risk of strangulation.
27. Do not service or maintain device and cu while in use.
APP
APP
APP
APP
120mmHg
120mmHg
USB
PC
BP Manager
Electromagnetic Compatibility Information
1. This device needs to be installed and put into service in accordance with the information provided
in the user manual.
2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the X5,
including cables specied by the manufacturer. Otherwise, degradation of the performance of this
device could result.
If higher IMMUNITY TEST LEVELS than those specied in Table 9 are used, the minimum separation
distance may be lowered. Lower minimum separation distances shall be calculated using the
equation specied in 8.10.
Manufacturers declaration-electromagnetic immunity
The X5 is intended for use in the electromagnetic environment specied below. The customer or
the user of the X5 should assure that is used in such and environment.
Immunity
test
IEC 60601 test level Compliance level Electromagnetic environment-
guidance
Conducted
RF IEC
61000-4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio bands
between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio bands
between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
X5 including cables, than the rec-
ommended separation distance
calculated from the equation ap-
plicable to the frequency of the
transmitter.
Recommended separation dis-
tance:
d = 1,2 √P, d = 1,2 √P 80MHz to 800
MHz, d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation dis-
tance in metres (m).
Interference may occur in the vi-
cinity of equipment marked with
the following symbol:
Radiated RF
IEC 61000-
4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is aected
by absorption and reection from structures, objects and people.
More information on EMC compliance of the device may be obtained from Rossmax using the
contacts shown in this manual.
IN0X50000000000XX
OBM_IB_X5(2)NA11_
BT_EN_SH_ver1823
WARNING: The symbol on this product means that it's an electronic product and following the European directive
2012/19/EU the electronic products have to be disposed on your local recycling centre for safe treatment.
Model: X5
www.rossmax.com
Warranty Card
This instrument is covered by a 5 year guarantee from the date of purchase. The guar-
antee is valid only on presentation of the warranty card completed or stamped by the
seller/dealer confirming date of purchase or the receipt. Batteries, cuff and accessories
are not included. Opening or altering the instrument invalidates the guarantee. The
guarantee does not cover damage, accidents or non-compliance with the instruction
manual. Please contact your local seller/dealer or www.rossmax.com.
Customer Name: ____________________________________________________
Address: ___________________________________________________________
Telephone: __________________________________________________________
E-mail address: ______________________________________________________
Product Information
Date of purchase: ___________________________________________________
Store where purchased:
____________________________________________________________________
Blood Pressure Monitor
Data Transfer via
Please download and install the
free APP onto your smartphone
- The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of such marks by Rossmax International Ltd. is under
license. Other trademarks and trade names are those of their respective owners.
- The blood pressure monitor uses Bluetooth®(Bluetooth® low energy technology)
- Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc.
- Google Play and the Google Play logo are trademarks of Google Inc.
Rossmax (Shanghai) Incorporation Ltd.
No. 6018, Huyi Rd., Waigang, Jiading District, Shanghai, China
Rossmax Swiss GmbH
Tramstrasse 16, CH-9442 Berneck, Switzerland
1

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Rossmax X5 BT wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Andere Handbücher von Rossmax X5 BT

Rossmax X5 BT Bedienungsanleitung - Holländisch - 2 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info