INTRODUCTION
The Fysic FB-100 blood pressure moni-
tor is developed to easily and precisely
measure your blood pressure.
By regularly checking your blood pres-
sure at home, you can provide your doc-
tor with valuable information about your
actual everyday blood pressure.
DECLARATION OF CONFORMITY
This product meets the requirements of
EC Directive 93/42/EEC
(Medical Equipment Directive).
The declaration of conformity is avail-
able on the website WWW.Fysic.NL
Do not simply throw away empty
or exhausted batteries, always
hand them over to your local
depot for Small Chemical Waste.
Power adaptor:
The back of this blood pressure monitor
is equipped with a connection for a power
adaptor with the following specifications:
input: 230V AC 50Hz
Output 6V DC 600mA
(attention: the outside is the + !!)
A power adaptor is not included.
CUFF
Connection to the blood
pressure monitor:
Connect the cuff to the
blood pressure monitor
according to the picture
below.
Attachment to the arm:
1. Lay down the cuff in front
of you as follows:
2. Remove the clothing or roll up the
sleeve to expose the left arm and
slide the cuff up around
the left arm until approx. 2
or 3 cm above the elbow.
The flexible tube should
come out of the cuff at the
inside of the elbow joint.
3. Wrap the cuff around the
arm and attach it to the
Velcro tape (don’t over
tighten).
MEASURING
Guidelines for accurate measure-
ments:
Always take measurements at the
same time of the day (preferably in
the morning) and under the same
conditions.
Never take measurements within half
an hour after drinking coffee or tea or
after smoking a cigar or cigarette.
Relax and sit down during measure-
ments, don’t speak or move. You
should relax your arm and put it in
front of you with the cuff at the height
of your heart.
Wait approx. 3 minutes before taking
the next measurement.
Blood circulation should
not be impeded, tight
clothing and other objects
should be removed.
Taking a measurement:
Attention: you can interrupt the measure-
ment at any moment by pressing the 0/I
button. The FB-100 will then switch itself
off instantly and the cuff will deflate.
1. Press the 0/I button to turn on the
FB-100.
2. The built-in pump will start to pressur-
ize the cuff. It’s possible for the pump
to start again once or twice to build
up some more pressure.
3. As soon as the desired pressure is
reached, the pump will stop while the
cuff gradually deflates;
while deflating, the blood pressure
and heart rate are measured
when the measurement is finished,
the display
will show
the blood
pressure
(both the systolic and diastolic
blood pressure) and the heart rate
while the cuff will fully deflate
the measured blood pressure and
heart rate are automatically stored
in the memory
4. The FB-100 will switch itself off auto-
matically after 3 minutes of you can
turn off the FB-100 quicker by press-
ing the 0/I button.
MEMORY
Each measurement is automatically
stored in the memory. The memory has
a capacity for 90 measurements. When
the memory is full, each new measure-
ment will erase the oldest stored mea-
surement.
Checking the memory:
Press the M. button to display the
average of the last 3 measurements.
Press the M. button again, ‘01’ will
appear followed by the results of the
last measurement.
Press the M. button again, ‘02’ will
appear followed by the results of the
second-to-last measurement.
Repeat this for all previous measure-
ments.
Erasing the memory:
Make sure to turn on the
FB-100.
Press and hold the M button
for approx. 5 seconds until
‘CLr appears in the display. All mea-
surements stored in the memory will
now be deleted.
USER’S MANUAL
FB-100
IMPORTANT
This product is intended only for
household use and for use by adults.
Keep the product and batteries out of
reach of children.
People suffering from cardiac
arrhythmia, vascular constriction,
arteriosclerosis (hardening of the
arteries) in the upper or lower limbs,
diabetes and heart patients with
pacemakers should consult their
doctor before measuring their blood
pressure, because in these cases
deviating blood pressure values
could be measured.
If you’re undergoing medical treat-
ment or if you’re using medicine, we
recommend you to always consult
your doctor when in doubt.
The use of this blood pressure
monitor should not be regarded as a
replacement for consulting your doc-
tor.
STANDARD VALUES
Depending on the person and situation,
a normal blood pressure is:
lower value (DIA): < 90
upper value (SYS): < 140
In case of ‘light hypertension’, the blood
pressure is:
lower value (DIA): 90-110
upper value (SYS): 140-180
In case of ‘serious hypertension’, the
blood pressure is:
lower value (DIA): >110
upper value (SYS): >180
During the entire day, our blood pres-
sure is continuously undergoing chang-
es. In the morning it rises heavily and
at the end of the morning it goes down
again. In the late afternoon our blood
pressure rises yet again, while it settles
on a low level at night. This is why your
measurements can vary when taken at
different times of the day.
POWER SUPPLY
Batteries:
1. Remove the battery compartment
cover at the bottom of the blood pres-
sure monitor.
2. Insert 4 AA 1.5V batteries according
to the correct polarity (see markings
inside the battery compartment).
Due to insufficient voltage, never
install rechargeable batteries.
3. Place the battery compartment cover
back onto the blood pressure monitor.
Batteries are not included
Change the batteries as soon as the
symbol
appears in the display of
when pressing the 0/I button has no
effect. Always replace all batteries at
once.
We recommend you to remove the bat-
teries from the blood pressure monitor
when not using the FB-100 for a long
period of time (>2 months).
Systolic pressure
Diastolic pressure
Heart rate
1

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Inhalt der Seiten


  • Page 1

    USER’S MANUAL

    INTRODUCTION

    STANDARD VALUES

    The Fysic FB-100 blood pressure monitor is developed to easily and precisely
    measure your blood pressure.

    Depending on the person and situation,
    a normal blood pressure is:
    • lower value (DIA): < 90
    • upper value (SYS): < 140

    By regularly checking your blood pressure at home, you can provide your doctor with valuable information about your
    actual everyday blood pressure.
    DECLARATION OF CONFORMITY
    This product meets the requirements of
    EC Directive 93/42/EEC
    (Medical Equipment Directive).

    The declaration of conformity is available on the website WWW.Fysic.NL

    FB-100

    In case of ‘light hypertension’, the blood
    pressure is:
    • lower value (DIA): 90-110
    • upper value (SYS): 140-180
    In case of ‘serious hypertension’, the
    blood pressure is:
    • lower value (DIA): >110
    • upper value (SYS): >180
    During the entire day, our blood pressure is continuously undergoing changes. In the morning it rises heavily and
    at the end of the morning it goes down
    again. In the late afternoon our blood
    pressure rises yet again, while it settles
    on a low level at night. This is why your
    measurements can vary when taken at
    different times of the day.

    -2-

    Do not simply throw away empty
    or exhausted batteries, always
    hand them over to your local
    depot for Small Chemical Waste.
    Power adaptor:
    The back of this blood pressure monitor
    is equipped with a connection for a power
    adaptor with the following specifications:
    input: 230V AC 50Hz
    Output 6V DC 600mA
    (attention: the outside is the + !!)
    A power adaptor is not included.
    CUFF
    Connection to the blood
    pressure monitor:
    Connect the cuff to the
    blood pressure monitor
    according to the picture
    below.
    Attachment to the arm:
    1. Lay down the cuff in front
    of you as follows:
    -6-

    2. Remove the clothing or roll up the
    sleeve to expose the left arm and
    slide the cuff up around
    the left arm until approx. 2
    or 3 cm above the elbow.
    The flexible tube should
    come out of the cuff at the
    inside of the elbow joint.
    3. Wrap the cuff around the
    arm and attach it to the
    Velcro tape (don’t over
    tighten).
    MEASURING
    Guidelines for accurate measurements:
    • Always take measurements at the
    same time of the day (preferably in
    the morning) and under the same
    conditions.
    • Never take measurements within half
    an hour after drinking coffee or tea or
    after smoking a cigar or cigarette.
    • Relax and sit down during measurements, don’t speak or move. You
    should relax your arm and put it in
    -7-

    -3-




    front of you with the cuff at the height
    of your heart.
    Wait approx. 3 minutes before taking
    the next measurement.
    Blood circulation should
    not be impeded, tight
    clothing and other objects
    should be removed.

    Taking a measurement:
    Attention: you can interrupt the measurement at any moment by pressing the 0/I
    button. The FB-100 will then switch itself
    off instantly and the cuff will deflate.
    1. Press the 0/I button to turn on the
    FB-100.
    2. The built-in pump will start to pressurize the cuff. It’s possible for the pump
    to start again once or twice to build
    up some more pressure.
    3. As soon as the desired pressure is
    reached, the pump will stop while the
    cuff gradually deflates;
    • while deflating, the blood pressure
    and heart rate are measured

    -8-

    IMPORTANT








    This product is intended only for
    household use and for use by adults.
    Keep the product and batteries out of
    reach of children.
    People suffering from cardiac
    arrhythmia, vascular constriction,
    arteriosclerosis (hardening of the
    arteries) in the upper or lower limbs,
    diabetes and heart patients with
    pacemakers should consult their
    doctor before measuring their blood
    pressure, because in these cases
    deviating blood pressure values
    could be measured.
    If you’re undergoing medical treatment or if you’re using medicine, we
    recommend you to always consult
    your doctor when in doubt.
    The use of this blood pressure
    monitor should not be regarded as a
    replacement for consulting your doctor.

    -4-

    • when the measurement is finished,
    the display
    Systolic pressure
    will show
    Diastolic pressure
    the blood
    Heart rate
    pressure
    (both the systolic and diastolic
    blood pressure) and the heart rate
    while the cuff will fully deflate
    • the measured blood pressure and
    heart rate are automatically stored
    in the memory
    4. The FB-100 will switch itself off automatically after 3 minutes of you can
    turn off the FB-100 quicker by pressing the 0/I button.
    MEMORY
    Each measurement is automatically
    stored in the memory. The memory has
    a capacity for 90 measurements. When
    the memory is full, each new measurement will erase the oldest stored measurement.

    -9-

    POWER SUPPLY

    Batteries:
    1. Remove the battery compartment
    cover at the bottom of the blood pressure monitor.
    2. Insert 4 AA 1.5V batteries according
    to the correct polarity (see markings
    inside the battery compartment).
    Due to insufficient voltage, never
    install rechargeable batteries.
    3. Place the battery compartment cover
    back onto the blood pressure monitor.
    Batteries are not included
    Change the batteries as soon as the
    appears in the display of
    symbol
    when pressing the 0/I button has no
    effect. Always replace all batteries at
    once.
    We recommend you to remove the batteries from the blood pressure monitor
    when not using the FB-100 for a long
    period of time (>2 months).
    -5-

    Checking the memory:
    • Press the M. button to display the
    average of the last 3 measurements.
    • Press the M. button again, ‘01’ will
    appear followed by the results of the
    last measurement.
    • Press the M. button again, ‘02’ will
    appear followed by the results of the
    second-to-last measurement.
    • Repeat this for all previous measurements.
    Erasing the memory:
    • Make sure to turn on the
    FB-100.
    • Press and hold the M button
    for approx. 5 seconds until
    ‘CLr’ appears in the display. All measurements stored in the memory will
    now be deleted.

    -10-



  • Page 2

    POSSIBLE PROBLEMS AND
    SOLUTIONSGELIJKE
    No reaction when turning on the
    FB-100:
    • Are the batteries exhausted or
    installed incorrectly? Replace them
    or insert correctly.
    “Err” appears in the display:
    • Impossible to pressurize the cuff.
    Check whether it’s loose and check
    the connection between the cuff and
    blood pressure monitor.
    • The air escapes too quickly. If this
    happens repeatedly, please contact
    the service department of Fysic.
    • Movement detected during the measurement. Please don’t move during
    the measurement.
    • Interfering signals near the FB-100.
    Remove any sources of interference,
    such as mobile phones.
    • Incorrect measurement value. The
    cuff is incorrectly attached. Please
    attach the cuff again.
    -11-

    SPECIFICATIONS

    • The FB-100 is used incorrectly, for
    example on a child.
    The FB-100 is only suitable for measuring the blood pressure of adults.

    Model:
    Fysic FB-100

    If it’s still not possible to measure the
    blood pressure, please contact the service department of Fysic. Never disassemble the blood pressure monitor in
    any circumstances.

    Manufacturer:
    Honsun (Nantong) Co., Ltd
    Address: No 8, Tongxing Road,
    Economic & Technical Development
    Area, Nantong City, Jiangsu, P.R.CHINA

    Keep in mind that not being able to take
    measurements can also be caused by
    the physical condition of the person. In
    this case, please consult your doctors.

    EU REPRESENTATION:
    SHANGHAI INTERNATIONAL
    HOLDING CORP. GmbH (Europe)
    Address: Eiffestrasse 80, 20537
    Hamburg, Germany






    Operating principle:
    Oscillometric measurement method
    Measurement range:
    20 – 280 mm Hg (blood pressure)
    40 – 195 pulse/min (heart rate)

    -12-

    STORAGE AND CLEANING

    WARRANTY

    Clean the outside of the unit with a moist,
    soft cloth.
    Stains on the cuff can be carefully
    removed with a moist cloth. The cuff may
    not be washed nor cleaned chemically.
    Store the product on a place not exposed
    to direct sunlight or high humidity.

    For the Fysic FB-100 Blood Pressure
    Monitor you have a warranty of
    24 months from the date of purchase.
    During this period, we guarantee the free
    repair of defects caused by material and
    workmanship errors. All this subjected to
    the final assessment of the importer.
    HOW TO HANDLE: If you notice any
    defect, first refer to the user’s manual.
    Please contact the supplier of this blood
    pressure monitor or contact customer
    service of Fysic via the Internet
    www.Fysic.nl

    -13-

    Accuracy:
    ± 3 mm Hg (cuff pressure)
    ± 5% (heart beat)
    (depending on temperature and
    humidity)

    Dimensions:
    120mm (d) x 85mm (w) x 65mm (h)
    Weight:
    220gr (excl. batt, excl. cuff)

    Cuff:
    arm circumference: 22-32cm
    Power supply:
    Battery 4x 1.5V battery, AA, Alkaline
    Adaptor DC 6V 600mA

    Protect against moisture
    Type BF

    Battery life:
    Up to approx. 300 measurements
    Ambient temperature, humidity and
    air pressure:
    Use:
    +10°C to +40°C

    <85% humidity

    700hPa - 1060hPa
    Storage: -20°C to +50°C

    <85% humidity

    500hPa - 1060hPa

    -14-

    -15-

    In case of unauthorised modifications
    and/or repairs performed by third parties.
    In case of incorrect transportation of the
    device without adequate packaging and
    when the device is not accompanied by
    this warranty card and proof of purchase.

    Service

    Help

    WWW.FYSIC.COM
    SERVICE@FYSIC.NL

    All further liability, especially concerning
    any consequential damages, is
    excluded.

    NL 073 6411 355
    (Lokaal tarief)

    BE 03 238 5666
    (Lokaal tarief)
    (Tarif local)

    DE 0180 503 0085
    Lokale Festnetzkosten

    Hesdo, Australiëlaan 1
    5232 BB, ‘s-Hertogenbosch

    THE WARRANTY BECOMES NULL
    AND VOID: In case of improper use,
    incorrect connections, leaking and/
    or incorrectly installed batteries, use
    of unoriginal parts or accessories,
    negligence and in case of defects
    caused by moisture, fire, flooding,
    lightning and natural disasters.

    The Netherlands

    Service

    WWW.FYSIC.COM
    SERVICE@FYSIC.NL

    NL 073 6411 355 (Lokaal tarief)
    (Lokaal tarief)

    -16-

    -17-

    -18-

    -19-

    Help

    BE 03 238 5666 (Tarif local)

    v1.0

    DE 0180-20503 0085 (Lokale Festnetzkosten)
    Hesdo, Australiëlaan 1, 5232 BB,
    ‘s-Hertogenbosch, The Netherlands






Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Fysic FB-100 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Holen Sie sich Ihr Benutzerhandbuch per E-Mail

Geben Sie Ihre E-Mail-Adresse ein, um das Handbuch zu erhalten von Fysic FB-100 in der Sprache / Sprachen: Englisch als Anhang in Ihrer E-Mail.

Das Handbuch ist 0,73 mb groß.

 

Sie erhalten das Handbuch in Ihrer E-Mail innerhalb von Minuten. Wenn Sie keine E-Mail erhalten haben, haben Sie wahrscheinlich die falsche E-Mail-Adresse eingegeben oder Ihre Mailbox ist zu voll. Darüber hinaus kann es sein, dass Ihr ISP eine maximale Größe für E-Mails empfangen kann.

Andere Handbücher von Fysic FB-100

Fysic FB-100 Bedienungsanleitung - Deutsch - 2 seiten

Fysic FB-100 Bedienungsanleitung - Holländisch - 2 seiten

Fysic FB-100 Bedienungsanleitung - Französisch - 2 seiten


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info