Introduction
Thank you for purchasing the unique 3in1 Compressor Nebulizer NK1000 which integrates with a
nebulizer kit, a nasal washer, and a nasal aspirator.
This product is developed in conjunction with respiratory therapists for the eective treatments
of asthma, COPD, chronic bronchitis, nasal and/or sinus congestion/stuness/ allergic symptoms
and other respiratory ailments. Treatments are delivered quickly, safely and conveniently making
this unit ideal for all ages in health care structures like hospitals and home care. Your compressor
nebulizer should be used under the supervision of a licensed physician and/or a respiratory
therapist. Together with your physician and/or respiratory therapist, you can feel comfortable
and condent knowing that you are obtaining the most eective inhalation and nasal aspiration
treatments for your respiratory condition.
NOTE: This is a medical device. Operate the device only as instructed by your doctor and/or
respiratory therapist. Please consult with your physician and/or pharmacist to determine if your
prescription medication is approved for use with this nebulizer. For type, dose and regime of
medication follow the instructions of your doctor or respiratory therapist.
Please read this manual carefully before use and be sure to keep this manual.
How the nebulizer kit works
Valve Adjustable Technology
The proprietary adjustable valve is able to deliver medications of
dierent viscosity level according to every users conditions and
needs at ease. Our VA technology allows users to adjust dierent
levels of nebulization rate ranging ≥ 0.1 – 0.3 ml/min at consistent
particle size.
The nebulization rates can be adjusted by the user in a very easy
way without exchanging parts. Higher nebulization rate/fully open
is for higher viscosity medications and higher breathing capacity
user while lower nebulization rate with closed valve will be more
appropriate for kids / infants with lower breathing capacity.
How the nasal washer works
Nasal washer connected to 3 in 1 compressor nebulizers, nebulizes an aqueous solutions for
ushing the nasal cavities by pressing the button on the ergonomic handle to activate the
nebulization and collects the return solutions. It can be used with saline solutions (isotonic and
hypertonic solutions, warm water) or medicated solutions for aerosol therapy. The body of nasal
washer has two adjacent chambers, inside one for holding the solution to be nebulized and the
outside for collecting the return solutions.
How the nasal aspirator works
The compressed air is generated by the suction pump which creates a vacuum sucking the mucus
into the collection chamber. Thanks to the Dual pump design, nebulization and suction pump,
provides the maximum hygiene to avoid cross infection and extends the life of the unit.
Preliminary Remarks
This device fulls the provision of the European Medical Device Directive 93/42/EEC amended
by 2007/47/EC and the European Standard EN 13544-1:2007+A1:2009 Respiratory therapy
equipment - Part 1: Nebulizing systems and their components.
Cautions
Please use general safety precautions when operating your nebulizer. This unit should be used
only for its intended purpose as described in this guidebook and with medications only under
the supervision and instruction of your physician. Do not use the device in anesthetic or ventilator
breathing circuits.
Product cautions
READ THE FOLLOWING BEFORE USING
To avoid electrical shock: keep unit away from water.
• Do not immerse the unit in liquid.
• Do not use while bathing.
• Do not reach for a unit that has fallen into water immediately unplug the unit.
Do not use the unit if it has any damaged parts (including plug), if it has been submersed in
water or dropped. Promptly send the unit for examination and repair.
The unit should not be used where ammable gas, oxygen or aerosol spray products are being
used.
Keep the air vents open. Do not place the unit on a soft surface where the openings can be
blocked.
• If any abnormality occurs, discontinue using until the unit has been examined and repaired.
The unit should not be left unattended while plugged in.
• Do not tilt or shake the unit when in operation.
• Disconnect the unit from the electrical outlet before cleaning, lling and after each use.
• Do not use attachments unless recommended by the manufacturer.
• Do not plug or unplug the AC adapter into the electrical outlet with wet hands.
• Unplug the AC adapter from the electrical outlet after using the device.
• Do not disassemble or attempt to repair the unit.
• Do not use the device in anesthetic or ventilator breathing circuits.
• Use only original Rossmax accessories/parts as included in the product.
Plug the device into the appropriate voltage outlet. Do not overload power outlets or use
extension power cords.
Operating cautions
Microbial contamination: personal use of all parts is recommended in order to prevent any risks
of contagious infection.
• Before each use, inspect the device and all parts. If damaged or deteriorated, replace it at once.
• Close adult supervision is highly recommended when the unit is used by children or invalids.
• Keep your eyes away from the output of medication mist.
• Do not use this unit while operating a vehicle.
• If any discomfort or abnormality occurs, stop using the unit immediately.
• Do not use the device if the air tube is bent.
• Do not service or maintain device while in use.
Do not use the tubing and/or AC adaptor for any other purpose than those specied, as they can
cause risk of strangulation.
• Manufacturer will not be held responsible for inappropriate use of the device.
The correct functioning of the device may be aected by electromagnetic interference caused
by malfunctioning television sets, radios, etc. If this occurs, move the source of interference away
from the device or plug it into a dierent power socket.
• Do not insert any object into the compressor.
Storage cautions
• Do not store the unit in direct sunlight, high temperature or humidity.
• Always keep the unit unplugged while not in use.
Some parts of the unit are small enough to be swallowed by children and can cause suocation;
keep the unit out of the reach of children.
• Store the device and parts in a clean location to avoid infection.
Cleaning cautions
Check air lter, nebulizer, mouthpiece and any other optional component before each use. Dirty
or worn parts should be replaced.
• Do not immerse the unit in water. It may damage the unit.
• Disconnect the unit from the electrical outlet before cleaning.
• Clean all necessary parts after each use as instructed in this guidebook.
Always dispose of any remaining medication in the medication cup after each use. Use fresh
medication each time you use the device.
Do not store the air tube with moisture or medication remaining in the air tube. This could result
in infection as a result of bacteria.
To prevent possible risk of infection from contaminated cleaning/disinfection solutions, always
prepare fresh solution for each cleaning cycle and discard solution after each use.
• Contaminated devices must be thoroughly sanitized prior to disinfection.
Do not use gravity autoclave, EOG gas disinfection or low temperature plasma sterilizer to disinfect
the unit and parts.
When disinfecting the unit by boiling, make sure to disassemble the parts and boil them with
plenty of water to completely immerse each individual part, taking care to avoid the parts directly
coming into contact with the bottom of the pot during the boiling.
Carry out the cleaning procedures after every application for maximum hygiene to avoid cross
infection and extend the life of the unit.
• Do not use any hard object to clean and/or detach the unit in order to avoid any damage.
• Do not use a microwave oven, dish dryer or hair dryer to dry the device or the parts.
Nebulizer kit: Usage
Pentamidine is not an approved medication for use with this device.
• If the medication cup is empty, do not attempt to operate the unit.
The maximum capacity of the medication cup is 5 ml and should not be overlled.
This product should not be used on patients, who are unconscious, sleeping, drowsy or are not
breathing spontaneously.
To avoid medication residue on the face, be sure to clean the patients face after removing the
mask.
Nasal washer: Usage
If the unit is used for therapeutic ushing, the administration of medication must be prescribed
or recommended by your physician, respecting the type of medication, doses and duration of
treatment indicated.
Nasal washer should be used with a physiological saline solution. Tap water” is not
recommended.
• Do not use the nasal washer with pneumatic device when taking a bath or shower.
• Do not add more than 20 ml of saline solutions to uid chamber.
• Do not insert any object into the nasal washer-tube connector.
Nasal aspirator: Usage
The nasal aspirator should be used exclusively for removing the nasal mucus. It cannot be used for
any other purpose.
• Do not obstruct the openings on the bottom and/or the side of the nasal aspirator.
If you notice rashes or redness appearing, inside or around the nose during or after using the unit
or anything unusual during use, stop using the unit and consult a physician.
• Do not attach the nasal aspirator to any other device.
MEDICAL DISCLAIMER:
This manual and product are not meant to be a substitute for advice provided by your doctor or
other medical professionals.
Don’t use the information contained herein or this product for diagnosing or treating a health
problem or prescribing any medication. If you have or suspect that you have a medical problem,
promptly consult your doctor.
Durable periods are as follows, provided the product is used to nebulize 2ml of medication 2 times
a day for 8 minutes each time, 15ml of saline solution 2 times a day for 5 minutes each time, and
aspirate 6 times a day for 3 minutes each time at room temperature (25°C).
Durable period may vary depending on usage and environment.
• Main unit 5 years
• Air Tube, Mouthpiece 1 year
• Nebulizer Kit 1 year
• Air Filter 60 days
• Nasal washer 1 year
• Adult and child masks 1 year
• Nasal aspirator 1 year
Product specications
AC Adaptor: DC12V, 1A
Power: AC100~240V, 50~60Hz
Power Consumption: ≤ 12 W
Sound Level: ≤ 60 dBA (1 meter away from NK1000 with Nebu-
lizer kit, Valve closed)
Compressor Pressure Range: ≥ 16 psi (110 kPa)
Operating Pressure Range: Nebulizer kit: ≥ 8 psi (55 kPa)
Nasal washer: ≥ 9 psi
Operating Vacuum Range: Nasal aspirator: ≤ -450mmHg
Operating Flow Range: Nebulizer kit: ≥ 3.0 lpm
Nasal washer: ≥ 4.2 lpm (for reference only)
Nasal aspirator: ≥ 5.0 lpm
Operating Temperature Range: 50ºF to 104ºF (10ºC to 40ºC)
Operating Humidity Range: ≤ 90% RH
Operating Atmospheric Pressure Range: 700-1060 hPa
Storage Temperature Range: -4ºF to 140ºF (-20ºC to 60ºC)
Storage Humidity Range: ≤ 90 % RH
Dimension (L x W x H): 260mmx170mmx120mm (10.24x6.69x4.72 inches)
Weight: 1540g (without accessories)
Medication Capacity: 5ml(cc) (Nebulizer kit)
Fluid chamber capacity: Max. 20 ml (c.c.) (Nasal washer)
Nasal adaptor with silicone tip size: L: 9mm, M: 7.6mm, S: 6.5mm (OD)
Particle Size (MMAD): ≤ 2.5μm (Nebulizer kit)
Average Nebulization Rate Fully open Valve: ≥ 0.3ml/min (Nebulizer kit) (0.9% Saline Solution)
Average Nebulization Rate Closed Valve: ≥0.1ml/min (Nebulizer kit) (0.9% Saline Solution)
Average Nebulization Rate: ≥ 3.0ml/min (Nebulizer washer) (0.9% Saline Solu-
tion)
Standard Accessories: Nebulizer Kit, Nasal Washer, Nasal Aspirator, Clear
Air Tube, Blue Air Tube, Mouthpiece, Filters (5pcs),
Adult and child masks, Filter (1pc)(Nasal Aspirator).
* Subject to technical modication without prior notice.
* Performance may vary with drugs such as suspensions or high viscosity, See drug supplier’s
data sheet for further details.
AProduct identication
1. Nebulizer Kit 2. Nasal Washer 3. Nasal Aspirator 4. Air Filter
5. AC Adaptor 6. Clear Air Tube 7. Blue Air Tube 8. Air Outlet
9. Vacuum regulator 10. Air Inlet 11. Power Switch 12. Air Filter Slot
13. AC Adaptor Jack
BUsing the AC Adapter
1. Check that the power switch is in the “OFF” position.
2. Connect the AC adapter with the AC adapter jack on the left side of the unit.
3. Plug the AC adapter into the socket. (AC adapters with required voltage and current indicated
near the AC adapter jack.)
4. Press the power-switch to the “ON” position.
C-2 Assembling your nebulizer kit
1. Gently twist and pull straight up on the lid of the nebulizer kit to separate into two parts
(medication cup and cover).
2. Make sure that the nozzle is properly installed on the upper cover. The stem inside the medication
cup inserts into the tube of the nozzle.
3. Add the prescribed amount of medication to the medication cup.
4. Reassemble the nebulizer kit by carefully twisting the medication cup and cover together. Make
sure that the two parts t securely.
5. Connect one end of the air tube in clear to the AIR OUTLET” connector marked “Nebulizer”
on the left side of the compressor and the other end to the air tube connector at the base of
the nebulizer kit.
C-3 Operating your nebulizer kit
The nebulizer kit is operable at up to a 45° angle.
If the angle is greater than 45°, no aerosol will
be generated.
1. Attach the angled mouthpiece or mask to
the top of the nebulizer kit.
2. The capacity of the medication cup is 2~5
ml.
3. Place the power switch to the “Nebulizer
(−) position.
4. When the compressor is acted on,
nebulization begins and aerosol is generated.
Inhale the medication and exhale through
the nebulizer kit.
DNasal washer
D-1 Product identication
1. Nasal adaptor with silicone tip (S/M/L) 2. Top cover 3. Separator 4. Nozzle
5. Fluid/ Nebulization chamber (Max. 20 ml/c.c.) 6. Return uid collection chamber
7. Ergonomic washer body 8. Air tube to the nebulizer
9. Nasal washer- tube connector 10. Trigger button
D-2 Assembling your nasal washer
1. Choose the most appropriate nasal adaptor with silicone tip (S/M/L) that ts the size of your
nostrils and insert it into the top cover.
2. Insert the nozzle into the center stem of the unit and make sure the nozzle is properly set in
place.
3. Pour the solution (10~15ml, max. 20 ml/c.c.) into the inside uid chamber. Always use fresh
solution each time you use the unit.
4. Put the separator and top cover back.
5. Connect one end of the air tube in clear to the AIR OUTLET connector marked “Washer” on
the left side of the compressor and the other end to the nasal washer- tube connector.
6. Make sure all parts are correctly inserted and connected to each other.
D-3 Operating your nasal washer
It is recommended to keep the nasal washer straight or operate at up to a 30° angle while in use.
If the angle is greater than 30°, the return uid from the nose may drip outside the unit.
1. Turn the power switch to the “Washer” (−) position.
2. Place the unit close to the nose and gently insert the nasal adaptor into one nostril.
3. During the application it is advisable to press your nger on the other nostril and keep it closed.
4. Start the nasal irrigator by holding down the trigger button while breathing in and release it
before breathing out. At this stage, to help expel the mucus, blow gently through the nostril
treated directly with the Nasal Washer.
5. Repeat above steps for the other nostril, trying to evenly distribute the solution between the
two nostrils. Interrupt the nebulization to blow your nose whenever necessary.
6. When used on children or disabled people, or in any case when there are problems in coordinating
the various steps as described 4, we recommend to run the application alternating the nostrils
about 20 seconds each, pausing in between to expel the mucus.
7. If the unit is used for therapeutic ushing, it is good idea to wash your nose with saline solution
before administering aerosol treatments. This will help to achieve a more eective distribution
of the medication over the nasal mucosa.
ENasal Aspirator
E-1 Product identication
1. Silicone tip 2. Silicone tube 3. Top cover 4. Finger opening
5. O-ring 6. Silicone tube 7. Collection chamber 8. Filter
E-2 Assembling your nasal aspirator
1. Insert the lter marked “IN” side to the air tube connector at the base of the nasal aspirator.
2. Make sure that the O-ring is properly attached.
3. Connect one end of the air tube in blue to the AIR INLET connector marked “Nasal
Aspirator” on the right side of the compressor and the other end to the air tube connector.
Make sure they are inserted rmly.
Instructions for use
Follow the cleaning instructions in this guidebook under “Cleaning procedures prior to using your
nebulizer for the rst time or after it has been stored for an extended period of time.
Note: Always unplug the compressor and make sure the power switch is turned to the “OFF” (O)
position before cleaning, assembling and before or after each use.
Operating your nebulizer
Place the compressor on a at, stable surface within reach. Make sure that the power switch is in
the OFF (O) position before use.
After every use:
1. Unplug the unit from the power source.
2. Allow the unit to completely cool.
3. Carefully detach the air tube from the nebulizer and pour out any remaining medication.
4. Follow the cleaning procedures provided in this guidebook.
Note: A 30-minute interval is recommended after each use.
CNebulizer kit
C-1 Product identication
1. Nebulizer kit 2. Nozzle 3. Angled Mouthpiece 4. Child Mask 5.Adult Mask
7
A
Max. 5ml (c.c.)
1
2
3
4
4
12
13
8
11
10
9
or
12 3
5
6
B
5
AC Adapter
(Ø3.8/Ø1.35)
C-1 C-2
C-3
<30°
D-3
Max.
20ml
(c.c.)
D-2
5
6
9
78
10
D-1
2
3
4
1
2

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Rossmax NK1000 3 in 1 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info